Oakton TDS 6+ handheld TDS meter with probe
Images are for representation only. Actual product may differ in size, color or options.
Oakton TDS 6+ Handheld Total Dissolved Solids Meter with Probe (each)
These microprocessor-based meters feature easy, push-button operation that offers high accuracy at an extremely affordable price. A quick and easy single-point calibration covers all five ranges with the TDS 6+ meters; for greater precision, single points can be calibrated within each of the five measurement ranges.
Additional features include a convenient Hold function and auto-off function that shuts meter off after 20 minutes of non-use. Programming functions allow customization of meter for temperature compensation, temperature coefficient, normalization, cell constant, and conductivity-to-TDS conversion factor for better accuracy in varying applications.
These microprocessor-based meters feature easy, push-button operation that offers high accuracy at an extremely affordable price. A quick and easy single-point calibration covers all five ranges with the TDS 6+ meters; for greater precision, single points can be calibrated within each of the five measurement ranges.
Additional features include a convenient Hold function and auto-off function that shuts meter off after 20 minutes of non-use. Programming functions allow customization of meter for temperature compensation, temperature coefficient, normalization, cell constant, and conductivity-to-TDS conversion factor for better accuracy in varying applications.
- Auto-ranging conductivity or TDS with temperature measurement
- TDS 6+ meters have five auto-ranging measurement modes — up to 200.0 ppt
- Temperature readout in °C with the push of a button
- Rubber boot with built-in stand protects meter and adds functionality
| Product Type: | TDS Meter |
| Brand: | Oakton |
| Range: | 0 to 10.00, 10 to 100.0, 100 to 1000 ppm; 1.00 to 10.00, 10.0 to 100.0, up to 200.0 ppt |
| Resolution: | 0.01, 0.1, 1 ppm; 0.01, 0.1 ppt |
| Accuracy: | ±1% full scale |
| Temperature Range: | -10 to 110°C |
| Temperature Resolution: | 0.1°C |
| Temperature Accuracy: | ±0.5° |
| ATC: | Yes (0.0 to 50.0°C) |
| Operating Modes: | TDS |
| Unit Size: | 5-1/2 x 2-3/4 x 1-5/16 inches |
| Includes: | Protective rubber boot with stand, conductivity/TDS/temperature probe (K = 1.0), and four AAA batteries. |
| Notes: | 5 maximum calibration points, one per conductivity range. Cell constant: 0.1, 1, and 10 (Selectable) |