Taylor Drop Test Kit, Peracetic Acid, Iodometric, 0-1000 ppm
Images are for representation only. Actual product may differ in size, color or options.
Taylor Drop Test Kit, Peracetic Acid, Iodometric, 0-1000 ppm
Peracetic acid (PAA) is a strong oxidizer and an excellent antimicrobial. It’s effective over a wide temperature range (32°–104°F/0°– 40°C); it has low potential for forming toxic byproducts, and when added to water it decomposes innocuously.
Although employed by industry for many years—for instance, in food and beverage production as a sanitizer and in pulp and paper manufacturing as a delignification and bleaching agent—peracetic acid has not been used to control the growth of microorganisms in cooling towers until recently. Now, since the U.S. Environmental Protection Agency has awarded registrations to biocides containing PAA that are intended to be kept at low levels, industrial water treaters have a need for a test in the range of 1–10 ppm.
As an alternative to an expensive and fragile test strip reader that requires periodic calibration, industrial customers may want to consider Taylor’s simple drop-count titration for monitoring PAA levels. Our peracetic acid test K-1546 will determine the concentration from 0–1000 ppm. The endpoint of the test is signaled by a distinct change in the sample from deep blue to no color at all, easy for anyone to see. Portable and able to withstand the rigors of field use, this kit will perform 110 tests on average.
Note: All oxidizers, including chlorine, are positive interferences for this test. Analysts should also be aware that a pH over 8, total hardness over 1000 ppm, sulfate over 1000 ppm, total alkalinity over 150 ppm, nitrite at any level, nitrate over 200 ppm, silica dioxide over 50 ppm, copper over 10 ppm, ferrous iron (Fe2+) at any level, and ferric iron (Fe3+) over 5 ppm can interfere with test results.
Peracetic acid (PAA) is a strong oxidizer and an excellent antimicrobial. It’s effective over a wide temperature range (32°–104°F/0°– 40°C); it has low potential for forming toxic byproducts, and when added to water it decomposes innocuously.
Although employed by industry for many years—for instance, in food and beverage production as a sanitizer and in pulp and paper manufacturing as a delignification and bleaching agent—peracetic acid has not been used to control the growth of microorganisms in cooling towers until recently. Now, since the U.S. Environmental Protection Agency has awarded registrations to biocides containing PAA that are intended to be kept at low levels, industrial water treaters have a need for a test in the range of 1–10 ppm.
As an alternative to an expensive and fragile test strip reader that requires periodic calibration, industrial customers may want to consider Taylor’s simple drop-count titration for monitoring PAA levels. Our peracetic acid test K-1546 will determine the concentration from 0–1000 ppm. The endpoint of the test is signaled by a distinct change in the sample from deep blue to no color at all, easy for anyone to see. Portable and able to withstand the rigors of field use, this kit will perform 110 tests on average.
Note: All oxidizers, including chlorine, are positive interferences for this test. Analysts should also be aware that a pH over 8, total hardness over 1000 ppm, sulfate over 1000 ppm, total alkalinity over 150 ppm, nitrite at any level, nitrate over 200 ppm, silica dioxide over 50 ppm, copper over 10 ppm, ferrous iron (Fe2+) at any level, and ferric iron (Fe3+) over 5 ppm can interfere with test results.
| Analyte | System | Method/Chemistry | Standard/Equivalence or Description | Comparator | Cell |
|---|---|---|---|---|---|
| Peracetic Acid | Drop test | Iodometric | 0-1000 ppm PAA | NA | 9198 |
| Product Type: | Test Kit |
| Brand: | Taylor Technologies |
| Product Line: | Drop Test |
| Class: | Peracetic Acid |
| Range: | 0-1000 ppm |
| Includes: | 2 of the 4030: PIPET,CAL(0.5&1.0 ML) W/ 1 of the 4078: PIPET,GRAD(3 ML) 1 of the 5015: INSTR,DT,PERACETIC ACID,D 1 of the 7060: CASE,DT,RIB,11.5X5.125X2, 1 of the 7265: LABEL,CASE,DT/SC,8.5X2.5 1 of the 8126: KIMPAC 4.75 X 11 1 of the 9198: SAMPLE TUBE,GRAD(25 ML) W 1 of the R-0925-C: PAA RGT #1,2 OZ 1 of the R-0926-A: PAA RGT #2,.75 OZ,DB 1 of the R-0927-C: PAA RGT #3,2 OZ 1 of the R-0928-C: PAA RGT #4,2 OZ 1 of the R-0929-C: PAA TITR SOL,2 OZ,DB |
| Notes: | Taylor items have a lead time of 3-15 days. |